From: Analytical validation of a multi-biomarker algorithmic test for prediction of progressive kidney function decline in patients with early-stage kidney disease
KIM-1
Repeatability
Sample
N
Mean Conc. (pg/mL)
SD
%CV
R1
80
3633
98.7
2.7
R2
1957
65.0
3.3
R3
1177
34.5
2.9
R4
669
22.8
3.4
R5
443
18.0
4.1
R6
109
3.0
R7
24
0.8
3.2