Skip to main content

Table 2 Repeatability estimates for sTNFR-1 based on a 20x2x2 study design (CLSI EP05-A3)

From: Analytical validation of a multi-biomarker algorithmic test for prediction of progressive kidney function decline in patients with early-stage kidney disease

sTNFR-1

Repeatability

Sample

N

Mean Conc. (pg/mL)

SD

%CV

R1

80

18,550

1062.5

5.7

R2

80

17,364

1420.8

8.2

R3

80

10,214

396.1

3.9

R4

80

4970

269.3

5.4

R5

80

3415

138.1

4.0

R6

80

2359

90.8

3.8

R7

80

1541

63.6

4.1