Skip to main content

Table 4 Within-laboratory precision summary for KIM-1 based on a multi-factor study design

From: Analytical validation of a multi-biomarker algorithmic test for prediction of progressive kidney function decline in patients with early-stage kidney disease

Sample

KIM-1

Within-laboratory

N

Mean Conc. (pg/mL)

SD

%CV

P1

100

3048

300.2

9.8

P2

100

1488

137.3

9.2

P3

100

752

72.6

9.7

P4

100

390

34.6

8.9

P5

100

212

19.0

8.9

P6

100

127

16.6

13.1

P7

100

77

8.2

10.7

P8

100

32

4.1

12.9