Table 6 Within-laboratory precision summary for sTNFR-2 based on a multi-factor study design

From: Analytical validation of a multi-biomarker algorithmic test for prediction of progressive kidney function decline in patients with early-stage kidney disease

Sample

sTNFR-2

Within-laboratory

N

Mean Conc. (pg/mL)

SD

%CV

P1

100

50,477

4175.7

8.3

P2

100

26,216

1889.1

7.2

P3

100

15,126

1125.0

7.4

P4

100

9732

719.0

7.4

P5

100

7177

528.4

7.4

P6

100

6107

592.1

9.7

P7

100

5216

387.0

7.4

P8

100

4551

355.1

7.8